COOK IRELAND LTD COLON DECOMPRESSION SET; FEG TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION Back to Search Results
Model Number G22181
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) # k171619.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
Colon decompression tube was placed in colon.Upon trying to withdrawal inner guiding catheter from outer drainage tube, resistance was felt and the inner guiding catheter broke in two pieces.One piece of broken inner guiding catheter came out of the outer drainage tube, while the other piece was stuck inside the distal end of the outer drainage tube.Colon decompression tube, with broken piece of inner guiding catheter, was then removed from patient.To finish the procedure, a new device of the same type was placed and no other issues were encountered.I recommended customer disassemble colon decompression set and flush and liberally lubricate each piece and then reassemble before placement.I also recommended using a shaking motion when withdrawing inner guiding catheter.Did any unintended section of the device remain inside the patient¿s body? no.If yes, please describe.Was the patient hospitalized or was there prolonged hospitalization due to this occurrence? no.If yes, please describe.Did the patient require any additional procedures due to this occurrence? no.If yes, please describe.Did the product cause or contribute to the need for additional procedures? no.If yes, please specify additional procedures and provide details.Has the complainant reported any adverse effects on the patient due to this occurrence? no.Has the complainant reported that the product caused or contributed to the adverse effects? no.Please specify adverse effects and provide details.Additional information provided by dm on 14feb2022: " is any patient information (age, weight, gender, pre-existing conditions, anatomical characteristics, etc.) available? not available was the patient hospitalized or was there prolonged hospitalization due to this occurrence? no.Did the patient require any additional procedures due to this occurrence? no.Did the product cause or contribute to the need for additional procedures? no.Has the complainant reported any adverse effects on the patient due to this occurrence? no.Has the complainant reported that the product caused or contributed to the adverse effects? no.Was the device flushed before use? yes.What was flushed through the device (water, saline, etc.)? water.Was lubrication applied to the decompression tube? yes.Details of the wire guide used (diameter, type, make)? used wire guide supplied with decompression tube.Please advise the anatomical location of the intended target site.Colon how experienced was the physician with using the cdsg? experienced was the secondary intervention performed during the same procedure as the device failure or was it scheduled for another day? same procedure were any other defects (other than the complaint issue) observed on the device prior to return (e.G.Kink)? no.What is the endoscope manufacturer and model number that was used? olympus 190.Was resistance encountered when advancing the decompression tube into position? unknown.If resistance was encountered, was the endoscope withdrawn a short distance? unknown.How often was the decompression tube irrigated? unknown.How long was the set left in dwelling? n/a"--did 15feb2022.For all complaints, ask: was the device flushed before use? n/a, yes, no what was flushed through the device (water, saline, etc.)? n/a, water, saline, other if other please specify.Was lubrication applied to the decompression tube? n/a yes, no.Details of the wire guide used (diameter, type, make)? please advise the anatomical location of the intended target site.How experienced was the physician with using the cdsg? did the patient require any additional procedures as a result of this event? n/a, yes, no.What intervention (if any) was required? was the secondary intervention performed during the same procedure as the device failure or was it scheduled for another day? n/a, same procedure, another day were any other defects (other than the complaint issue) observed on the device prior to return (e.G.Kink)? n/a, yes, no.If yes, please specify what was observed and where on the device it was observed.What is the endoscope manufacturer and model number that was used? was resistance encountered when advancing the decompression tube into position? n/a, yes, no.If resistance was encountered, was the endoscope withdrawn a short distance? n/a, yes, no.How often was the decompression tube irrigated? did any section of the device detach inside the endoscope or patient? n/a, yes, no.How long was the set left in dwelling?.
 
Additional Manufacturer Narrative
Pma/510(k) # k171619.Device evaluation: 1 x cdsg-14-175 device of lot number c1862939 involved in this complaint was not available for evaluation.Document review including ifu review: prior to distribution cdsg-14-175 devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for cdsg-14-175 of lot number c1862939 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1862939.The instructions for use states the following: ¿if package is opened or damaged when received, do not use.Visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use.¿ root cause review: a definite root cause cannot be determined as the device was not returned to cirl for evaluation.A possible root cause could be attributed to excessive force used when withdrawing the catheter from the drainage tube which contributed to the breakage that occurred.Summary: complaint is based on customer testimony.According to the information reported, there were no additional procedures required or adverse effects to the patient.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
Supplemental report is being submitted due to completion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLON DECOMPRESSION SET
Common Device Name
FEG TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
sinead o'leary
o halloran road
national technology park
limerick 
MDR Report Key13706025
Report Number3001845648-2022-00123
Device Sequence Number6476720
Product Code FEG
UDI-Device Identifier00827002221819
UDI-Public(01)00827002221819(17)240831(10)C1862939
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberG22181
Device Catalogue NumberCDSG-14-175
Device Lot NumberC1862939
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/09/2022
Event Location Hospital
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer02/14/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date11/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-