ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 32MM B; PROSTHESIS, HIP Back to Search Results
Catalog Number 010000924
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Osteopenia/ Osteoporosis (2651)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product is in the process of being returned to zimmer biomet for investigation.Once the investigation has been completed a follow-up mdr will be submitted report source: (b)(6).Multiple mdr reports were filed for this event, please see associated reports 0001825034 -2022 -00491.
 
Event or Problem Description
It was reported that after the intraoperative acetabular was reamed to a 45 and placing the 46mm g7 cup implant, the trial liner was put in.Then, the overall trial was completed and the angle was found to be suitable.Therefore, the 32/b high wall liner was opened as requested by the surgeon.The surgeon first cleaned the soft tissue around the cup edge, checked the acetabular base and then struck the liner.However, the g7 liner did not lock into place.The surgeon cleaned the soft tissue around the cup, the bone bulge, again checked the angle and struck the liner, but it still would not lock.The surgeon repeated checks and cleanups for up to 20 minutes, but the liner would not lock into place.Due to the patient's bone quality problem, the patient ended up with a collapsed acetabulum.A ceramic liner, a g7 cup, and the bottom of the acetabulum was filled with autologous bone and allogeneic bone to complete the procedure.As a result of the malfunction, the procedure was prolonged and caused bleeding to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.B1 correction: complaint was for product problem, no adverse event.One g7 hi-wall e1 liner 32mm b item# 010000924 lot# 6968592 was returned and evaluated.Upon visual inspection there was no visible damage to either the liner or the shell.The liner was sat in the shell and the scallops fit in the grooves of the shell.The devices were not assembled in the lab.A review of the device history records identified no related deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information on the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL E1 LINER 32MM B
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13709265
Report Number0001825034-2022-00483
Device Sequence Number15024599
Product Code LPH
UDI-Device Identifier00880304527003
UDI-Public(01)00880304527003(17)260211(10)6968592
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2026
Device Catalogue Number010000924
Device Lot Number6968592
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2022
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age77 YR
Patient SexFemale
Patient Weight53 KG
-
-