| Catalog Number |
301-05-01 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during impaction a little sliver of metal broke off on the left side of the broach handle between 0 and 10 degrees of version.The broach handle works but losing that piece of metal has created a sharp edge.No parts or pieces fell into the patient¿s wound site.There was no surgical delay.Patient was last known to be in stable condition following the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (d4) lot number: 264006002.(h3) the broken instrument reported was likely the result of the small bridge area thickness of 03.035 inches, coupled with movement of the pull release shaft (301-200-01) during implantation, which likely make the bridge area susceptible to failure.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.As determined, the broken instrument reported was likely the result of the small bridge area thickness of 0.035 inches, coupled with movement of the pull release shaft (301-200-01) during impaction, which likely made the bridge area susceptible to failure.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.As determined, the broken instrument reported was likely the result of the small bridge area thickness of 0.035 inches, coupled with movement of the pull release shaft (301-200-01) during impaction, which likely made the bridge area susceptible to failure.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were corrected: d1, h6.As determined, the broken instrument reported was likely the result of the small bridge area thickness of 0.035 inches, coupled with movement of the pull release shaft (301-200-01) during impaction, which likely made the bridge area susceptible to failure.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|