Model Number WNDARM
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Osteomyelitis (4533)
Event Date 02/05/2022
Event Type  Injury  
Event Description
On 07-feb-2022, the following information was reported to kci by the patient: on (b)(6) 2022, the patient was hospitalized allegedly for a bone infection and has covid.The patient will be going home and will be on 30 days of antibiotics.No additional information was provided.On 09-dec-2021, the device was tested per quality control procedure by a kci service center, and the device passed and met specification.On (b)(6) 2021, the device was placed with the patient.On 04-mar-2022, the device was tested per quality control procedure by a kci service center and the unit passed the quality control checks and met specifications.Inspection and testing of the device did not reveal any evidence of an operational malfunction with the unit.
 
Manufacturer Narrative
Hold for kp 3.14 based on the information provided, it cannot be determined that the alleged hospitalization due to bone infection requiring antibiotics is related to the activ.A.C.¿ ion progress¿ remote therapy monitoring system.Kci has made multiple unsuccessful attempts to obtain additional clinical information.The device passed quality control checks before and after placement.Device labeling, available in print and online, states: infected wounds should be monitored closely and may require more frequent dressing changes than non-infected wounds, dependent upon factors such as wound conditions and treatment goals.Refer to dressing application instructions (found in v.A.C.® dressing cartons) for details regarding dressing change frequency.As with any wound treatment, clinicians and patients/caregivers should frequently monitor the patient's wound, periwound tissue and exudate for signs of infection, worsening infection or other complications.Some signs of infection are fever, tenderness, redness, swelling, itching, rash, increased warmth in the wound or periwound area, purulent discharge or strong odor.Infection can be serious, and can lead to complications such as pain, discomfort, fever, gangrene, toxic shock, septic shock and/or fatal injury.Some signs or complications of systemic infection are nausea, vomiting, diarrhea, headache, dizziness, fainting, sore throat with swelling of the mucus membranes, disorientation, high fever, refractory and/or orthostatic hypotension or erythroderma (a sunburn-like rash).If there are any signs of the onset of systemic infection or advancing infection at the wound site, contact the treating physician immediately to determine if v.A.C.® therapy should be discontinued.Precautions: the v.A.C.® therapy system will not be effective in addressing complications associated with the following: ischemia to the incision or incision area, untreated or inadequately treated infection, inadequate hemostasis of the incision, cellulitis of the incision area.Osteomyelitis: v.A.C.® therapy should not be initiated on a wound with untreated osteomyelitis.Consideration should be given to thorough debridement of all necrotic, non-viable tissue, including infected bone (if necessary), and appropriate antibiotic therapy.Protect intact bone with a single layer of non-adherent material.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM
Type of Device
OMP
Manufacturer (Section D)
KINETIC CONCEPTS, INC.
san antonio TX 78249
Manufacturer Contact
steven jackson
6203 farinon drive
san antonio, TX 78249
2102556438
MDR Report Key13711839
MDR Text Key287279142
Report Number3009897021-2022-00049
Device Sequence Number1
Product Code OMP
UDI-Device Identifier00849554005600
UDI-Public0100849554005600
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberWNDARM
Device Catalogue Number420095
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Date Manufacturer Received02/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/14/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexMale
-
-