MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Battery Problem (2885); Connection Problem (2900); Material Integrity Problem (2978)
|
Patient Problems
Dysphagia/ Odynophagia (1815); Muscular Rigidity (1968); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/06/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the silver pin on the desktop charger broke.Troubleshooting was not required.The issue was not resolved through troubleshooting.Technical services sent a request to replace the desktop charger.Additional information received from the consumer who reported a potential emergency situation of return of symptoms (emergency because the patients become so frozen, they were like a rock and experienced swallowing issues) if the patient¿s neurostimulator charge would run out.The neurostimulator was checked with the implant at 40% but were having trouble swallowing and they expected the patient¿s charge to last until noon today.It was noted the consumer charged every day and a half because they used maximum settings.The consumer was working on providing emergency care in case it would be needed, but hoped they could get a desktop charger sooner.
|
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: 37761, serial#: (b)(4), product type: recharger.Other relevant device(s) are: product id: 37761, serial/lot #: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
"h3: analysis of the desktop charger serial #(b)(6) revealed the connector pin was broken." medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|