• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. EB EXC GRN 30IN 0 S/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. EB EXC GRN 30IN 0 S/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number X834H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a arteriovenous fistula procedure on (b)(6) 2022 and suture was used.During a skin bursa for closure of an entry point of an arteriovenous fistula, it was noted that the end of the needle broke while making a stitch.Another like device was used to complete the procedure.There were no patient consequences reported.This incident caused a risk of exposure to blood and a risk of losing a needle tip in the patient.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).Product complaint number: pc-001067747 a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 5/17/2022.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional h3 investigation summary: a failed suture needle was submitted fractographic evaluation.The component was identified as product code x834h.A fracture was observed at the suture attachment of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.A microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage observed coincidental to the fracture provides additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EB EXC GRN 30IN 0 S/A SH
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo PR
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13712126
Report Number2210968-2022-01592
Device Sequence Number20195742
Product Code GAT
UDI-Device Identifier10705031058064
UDI-Public10705031058064
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model NumberX834H
Device Catalogue NumberX834H
Device Lot NumberRHBADU
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer04/25/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-