• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL EPIC SUPRA VALVE W/FLEXFIT; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EPIC SUPRA VALVE W/FLEXFIT; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number ESP200-25
Medical Device Problem Codes Incomplete Coaptation (2507); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 02/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2022, a 25 mm epic plus supra valve was successfully implanted without issue.Intraoperative echo preformed revealed moderate aortic insufficiency from the device.The patient was placed back on bypass, the valve removed, and replaced with a 23 mm epic plus supra valve to resolve the event.There was no clear cause for the moderate aortic insufficiency.The procedure was extended longer than is clinically significant, however the patient remained hemodynamically stable throughout.The patient is stable and is recovering.No additional information was provided.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
An event of moderate aortic insufficiency was reported.A returned device assessment and histopathological examination could not be performed as the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was also reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.H6.Medical device problem code: 2507 incomplete coaptation replaced 3190 insufficient information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC SUPRA VALVE W/FLEXFIT
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL LTDA. REG#3001883144
1301rua profvieira demendonça
bairro engenho nogueira 31.31 0-26
BR   31.310-260
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13712329
Report Number2135147-2022-00074
Device Sequence Number5448543
Product Code LWR
UDI-Device Identifier05415067033147
UDI-Public05415067033147
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P040021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/19/2022
Device Model NumberESP200-25
Device Catalogue NumberESP200-25
Device Lot NumberBR00030416
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/11/2022
Supplement Date Received by Manufacturer07/18/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexUnknown
-
-