| Model Number |
COBAS 8000 CORE |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter questioned results for multiple patient samples tested for multiple assays on a cobas e801 module.The customer identified a clogged sipper nozzle.After cleaning the sipper nozzle, the customer repeated patient samples and received different results.Discrepant results were provided for 3 patient samples tested for elecsys anti-tg (anti-tg) and elecsys anti-tpo (anti-tpo).Patient 1 initial anti-tg result was 139 iu/ml.The repeat result was 58.9 iu/ml.Patient 2 initial anti-tpo result was 600 iu/ml.The repeat result was 305 iu/ml.Patient 3 initial anti-tg result was 64.5 iu/ml.The repeat result was 15.5 iu/ml.Patient 3 initial anti-tpo result was 72.6 iu/ml.The repeat result was 9.00 iu/ml with a data flag.The initial results were reported outside of the laboratory.The reagent lot numbers with expiration dates were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
After cleaning the pre-wash sipper nozzle, the customer had no further issues.The investigation determined the clogged pre-wash sipper nozzle was the cause of this event.
|
| |
|