• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. FIXED SCREW, SELF-STARTING; SIZE Ø4.35X18 MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. FIXED SCREW, SELF-STARTING; SIZE Ø4.35X18 MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Model Number 8801-14318DA
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an ozark fixed screw at c4 fractured post-operatively.Revision surgery will not be performed.
 
Additional Manufacturer Narrative
Device remains implanted in patient.
 
Additional Manufacturer Narrative
Visual, dimensional, material and functional analysis could not be performed as the screw associated with this record was not returned.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.Complaint history records were reviewed for this catalog number, no similar complaints were identified.Initial report states an ozark screw failed at c4, the top level of a corpectomy construct.It was reported that the patient will not be revised and that only one screw fractured.No trauma was experienced by the patient.No additional patient information was available.X-ray provided confirms that one cranial screw fractured.The x-ray provided was taken after event occurred.No post-op/pre-event x-ray was available to review.Cranial end of plate is not contacting the anterior surface of the vertebral body in the post-event x-ray.It cannot be confirmed if the plate was touching vertebral body after initial procedure.Ozark view and guide surgical technique was reviewed: ¿the plate should sit flush on the vertebral body.Excessive forces may be experienced by the screws if the plate is not in contact with the vertebral body¿.¿these internal fixation devices are load sharing devices which maintain alignment until healing occurs.If healing is delayed or does not occur, the implant could eventually break, bend, or loosen.Physical activity and load bearing have been implicated in premature loosening, bending or fracture of internal fixation devices in the absence of complete bone healing.¿ due to lack of product information and device not returned, an exact cause of the reported event of the screw fracture cannot be conclusively determined at this time.
 
Event or Problem Description
It was reported that an ozark fixed screw at c4 fractured post-operatively.Revision surgery will not be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIXED SCREW, SELF-STARTING; SIZE Ø4.35X18 MM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13712783
Report Number3004774118-2022-00094
Device Sequence Number17761522
Product Code KWQ
UDI-Device Identifier10888857392557
UDI-Public10888857392557
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K172104
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8801-14318DA
Device Catalogue Number8801-14318DA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/07/2021
Supplement Date Received by Manufacturer03/21/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-