MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems Pumping Stopped (1503); Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
Information was received from multiple sources (manufacturer representative (rep), healthcare provider (hcp)) regarding a patient who was receiving morphine (concentration of 25 at dose of 3. 0) via an implantable pump for non-malignant pain. It was reported that the pump battery alarmed and indicated stopped.  there were no known external factors that may have led or contributed to the issue. A dye study was done which revealed an issue near the pump connection. It was noted that the catheter was sutured therefore cutting off infusion flow and caused the pump to alarm. The hcp determined it was an issue from the original implant. The pump and catheter were replaced. The issue was resolved at the time of report. The patient's weight and medical history were asked but unknown. The patient's status at the time of report was alive - no injury.
 
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial#: (b)(4), implanted: (b)(6) 2016, explanted: (b)(6) 2022, product type: catheter. Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 15-oct-2017, udi#: (b)(4). If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13713041
MDR Text Key290067335
Report Number3004209178-2022-03050
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/14/2018
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2022
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured09/24/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/09/2022 Patient Sequence Number: 1
-
-