ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ZEPHYR SR; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5620
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented in clinic for follow up.The pacemaker showed premature battery depletion.The pacemaker was explanted and replaced.The patient was stable.
 
Additional Manufacturer Narrative
The reported event of ¿premature discharge of battery¿ could not be confirmed.The device was received at elective replacement indicator level.The device was then programed to nominal settings for the remainder of the analysis.Electrical testing indicated normal device characteristics.Battery assessment showed the device at elective replacement indicator level due to normal usage and autocapture being enabled.Longevity assessment showed normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR SR
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13713262
Report Number2017865-2022-05038
Device Sequence Number15072151
Product Code NVZ
UDI-Device Identifier05414734501606
UDI-Public05414734501606
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2017
Device Model Number5620
Device Catalogue Number5620
Device Lot NumberA000009856
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2022
Supplement Date Received by Manufacturer04/07/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date04/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/08/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-