| Catalog Number |
03.010.485 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, during an unknown procedure, when drilling through the oblique holes on the aiming arm, the drill did not align and hit the nail.The surgeon had to remove the drill sleeve and drill through outer protection sleeve.No further information was provided.This is report 1 of 6 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: the investigation could not be completed; no conclusion could be drawn, as no product was received.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot part number: 03.010.485 lot number: 9590916 part/lot combination is unknown at jabil (in epr sap p02), no dhr review possible for group "synthes trauma - mfg - dhr - hagendorf".Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|