Model Number AX083020003
Medical Device Problem Code Device Tipped Over (2589)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician arrived onsite to inspect the unit and found that the screws that attach the docking station to the floor were loose.As the screws were loose, this prevented the transfer carriage from locking properly to the docking station.The technician also found that the locking mechanism had been damaged as a result of the reported event.The unit has been removed from service pending repairs.A follow-up mdr will be submitted when additional information becomes available.
 
Event or Problem Description
The user facility reported that their evolution transfer carriage fell to the floor.No report of injury.
 
Additional Manufacturer Narrative
The technician made the necessary repairs, tested the unit, confirmed it to be operating according to specifications and returned it to service.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION TRANSFER CARRIAGE
Common Device Name
TRANSFER CARRIAGE
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key13715224
Report Number3005899764-2022-00007
Device Sequence Number17638810
Product Code FLE
UDI-Device Identifier00724995150389
UDI-Public00724995150389
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAX083020003
Device Catalogue NumberAX083020003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer02/07/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date03/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/14/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-