| Catalog Number |
2C8671 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred during the unspecified date of (b)(6) 2022 (2 weeks ago).Device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a clearlink system extension set was leaking coming from the filter area.This was observed an infusion of irinotecan to a patient in the cancer center.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: lot #: ¿r21i06086¿, previously ¿asku¿.D4: udi # ¿(01)00085412046334¿, previously submitted as ¿ni¿.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to specifications.A functional testing was performed including clear passage and pressure testing with no issues noted.Additional priming was performed with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|