| Model Number |
25195 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.Testing found the display has a led segment that does not power on due to defective component d2 on instrument board.A service history record review reveals that this unit was in the field for over three (3) years with no previous reported issues related to this reported event.
|
| |
|
Event or Problem Description
|
|
The customer reported the segments of the screen aren't lighting up.No patient impact or consequences were reported.
|
| |
|