Model Number PPLUS120
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a therapeutic peripheral procedure in a slightly calcified distal sfa and proximal popliteal, the manufacturer's catheter¿s needle did not deploy correctly.The needle was fired at 3 mm but could not be advanced to the true lumen.Therefore, the needle was retracted and the wire was pulled back in to adjust the needle depth to 5 mm, with no success.Then, the catheter was advanced forward 1 cm at a time, deployed the needle and repeated approximately 5 separate times, again with no success.Angiogram showed that the needle appeared to be pointing in a different direction; however, there was no evidence of needle penetrating the incorrect vessel wall.The physician decided to remove the catheter, but it got stuck on the non-manufacturer's wire, thus were removed together as a unit.The needle was fully retracted when removed from the patient.A new manufacturer's catheter and new non-manufacturer''s wire were used to complete the procedure.There was no patient injury reported.This product problem is being submitted because the manufacturer's catheter needle was not pointing in the desired direction.There is a potential for harm if the malfunction were to recur.
 
Additional Manufacturer Narrative
This complaint was reviewed and investigated according to the manufacturer¿s policy.The pioneer plus pplus120 was not returned for evaluation, thus no returned product investigation was performed.
 
Additional Manufacturer Narrative
Blocks d9 & g3: the pioneer plus catheter was returned for evaluation.Block h3: the pioneer plus catheter was visually and microscopically inspected.The tip of the catheter was deformed and a tear was noted at the exit port.During functional testing, the needle could not be deployed due to resistance.However, the needle was intact with no rough/sharp edges or missing material observed.Block h6: the probable cause of the reported "needle misaligned" could not be determined since the needle failed to deploy during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIONEER PLUS PPLUS120
Common Device Name
CATHETER FOR CROSSING TOTAL OCCLUSIONS
Manufacturer (Section D)
PHILIPS VOLCANO
3721 valley centre drive #500
san diego CA 92130
Manufacturer (Section G)
VOLCARICA S.R.L.
coyol free zone &business park
b37
alajuela
CS  
Manufacturer Contact
ayse yarimoglu
3721 valley centre drive #500
san diego, CA 92130
MDR Report Key13716068
Report Number3008363989-2022-00015
Device Sequence Number17713820
Product Code PDU
UDI-Device Identifier00845225002565
UDI-Public(01)00845225002565(11)210806(17)230806(10)5161258
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K162418
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/06/2023
Device Model NumberPPLUS120
Device Catalogue Number400-0200.246
Device Lot Number5161258
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ABBOTT: 0.014" SPARTACORE GUIDEWIRE; LAKE REGION MEDICAL: 0.014" PHOENIX LIGHT SUPPORT; UNK MFG AND SIZE: INTRODUCER SHEATH
Patient SexUnknown
-
-