| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: unknown tibia base plate.Report source: foreign country: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, due to the device was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the nexgen inlay could not be anchored to the tibia base plate.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: h2; h3; h6 reported event was unable to be confirmed due to limited information received from the customer.The device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|