| Model Number |
PVK-1070Z |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If additional information becomes available, a supplemental report will be filed with the new information.This device is classified as import for export, therefore 510k is not applicable.
|
| |
|
Event or Problem Description
|
|
No "knowledgement" of adverse event.Refer to expiration date as complaint notification.Customer advised: poor image quality.The time of event is unknown.This event occurred, no report of patient harm.The date of event is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information: g6: follow up #2, h2:if follow-up, what type? h3:device evaluated by manufacture.Additional information: h4:device manufacture date.Evaluation summary: it is presumed that the event was caused by a failure in the imaging system due to long-term use (9 years) and a decrease in imaging capability.(no repair history) as for the cause of the deterioration of the imaging capability, it is thought that there are factors such as failure of the image sensor, damage to the ccd cable, breakage of the circuit pattern, etc., but we have not reached a conclusion.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information: g6: follow up #2.H2:if follow-up, what type? model pvk-1070z is video adaptor module for fiberscope, so there is no 510k number.
|
| |
|