Model Number PVK-1070Z
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If additional information becomes available, a supplemental report will be filed with the new information.This device is classified as import for export, therefore 510k is not applicable.
 
Event or Problem Description
No "knowledgement" of adverse event.Refer to expiration date as complaint notification.Customer advised: poor image quality.The time of event is unknown.This event occurred, no report of patient harm.The date of event is unknown.
 
Additional Manufacturer Narrative
Correction information: g6: follow up #2, h2:if follow-up, what type? h3:device evaluated by manufacture.Additional information: h4:device manufacture date.Evaluation summary: it is presumed that the event was caused by a failure in the imaging system due to long-term use (9 years) and a decrease in imaging capability.(no repair history) as for the cause of the deterioration of the imaging capability, it is thought that there are factors such as failure of the image sensor, damage to the ccd cable, breakage of the circuit pattern, etc., but we have not reached a conclusion.
 
Additional Manufacturer Narrative
Correction information: g6: follow up #2.H2:if follow-up, what type? model pvk-1070z is video adaptor module for fiberscope, so there is no 510k number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
VIDEO MODULE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key13716707
Report Number9610877-2022-00361
Device Sequence Number17247129
Product Code FEA
UDI-Device Identifier04961333240364
UDI-Public04961333240364
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPVK-1070Z
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer02/07/2022
02/07/2022
Initial Report FDA Received Date03/09/2022
Supplement Report FDA Received Date12/02/2022
12/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-