• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 58MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 58MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-01-058
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Noise, Audible (3273)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Discomfort (2330); Joint Laxity (4526)
Date of Event 02/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon reported to bfarm: "hip-tep left side on (b)(6) 2021 in (b)(6).Patient now is transferred by ambulance from the (b)(6) clinic in (b)(6).Presentation there because of subluxation phenomena at the left hip joint.He had never been free of discomfort since the operation.Always when climbing stairs and during certain movements and loads, he had heard a creaking noise and had slight pain.So far, he has had the feeling that the hip joint has jumped out six times, but then located back in on its own.Yesterday renewed subluxation phenomenon when bending, this time with severe pain, again snapping back by itself.He now has a considerable feeling of instability and does not dare to make certain movements.No pain at rest.No trauma.Radiological dislocation of the pe.Revision of head and inlay on (b)(6) 2022 in (b)(6)." doi: (b)(6) 2021; dor: (b)(6) 2022; left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d1, d2, d4 (catalog), d10 and g1.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was not received for analysis.Examination of the returned liner involved in this complaint, could confirmed that the poly liner underwent a disassociation event from the cup.Interlock feature edge of the liner denoted three ard tabs sheared off from the liner which is evidence of the liner having disassociated from the cup.It is not unreasonable that noise would be present due to the ceramic head articulating against the metal cup.However, audible noise depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : a dhr was performed: 1) quantity manufactured: (b)(4) parts, 2) date of manufacture: 03/nov/2020, 3) any anomalies or deviations identified in dhr: none, 4) expiry date: 31/oct/2030, 5) ifu reference: 090200701.Device history review : a manufacturing record evaluation was performed for the finished device (121701058/9627500) product and lot numbers, and no non-conformances were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Mwr-17052022-0001185278 has been retracted is this was reported in error.The device has not been received and still in transit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE 100 ACET CUP 58MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork IN
EI  
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13719703
Report Number1818910-2022-04380
Device Sequence Number14966245
Product Code LPH
UDI-Device Identifier10603295008583
UDI-Public10603295008583
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-01-058
Device Catalogue Number121701058
Device Lot Number9627500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer03/21/2022
04/01/2022
04/01/2022
05/17/2022
06/21/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date03/31/2022
04/04/2022
05/17/2022
05/19/2022
06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; CORAIL2 NON COL HO SIZE 14; DELTA CER HEAD 12/14 32MM +1; PINN MAR NEUT 32IDX58OD; UNK HIP ACETABULAR LINER POLY; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP FEMORAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-