Model Number 5480
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During the previous follow-up, approximately seven months ago, the battery longevity showed less than two years until elective replacement indicator (eri).However, during the most recent follow-up, the battery was now at eri.Premature battery depletion was suspected.Device replacement is anticipated.The patient was stable.
 
Event or Problem Description
New information was received indicating the device was explanted and replaced.The patient was stable.
 
Additional Manufacturer Narrative
The reported event of premature battery depletion was not confirmed.The device was received with battery voltage at elective replacement level which was reached post-explant.Device image indicated that auto-capture was enabled and operated in high output mode at times consistent with the false replacement alert triggered in the field.When automatic settings are programmed, such as auto-capture, the device output would adjust itself to accommodate the patient¿s needs for pacing and thus affect the estimated longevity of the device.Analysis performed, in nominal settings, and at various outputs, did not find any anomalies.Longevity assessment was performed, and the device exceeded projected longevity based on field settings indicating normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDENTITY ADX VDR
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13720075
Report Number2017865-2022-04992
Device Sequence Number4807403
Product Code LWP
UDI-Device Identifier05414734006927
UDI-Public05414734006927
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2014
Device Model Number5480
Device Catalogue Number5480
Device Lot Number3835217
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer04/01/2022
06/23/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date04/08/2022
06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-