Model Number CH-S190-XZ-Q
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device was returned to olympus for evaluation, upon inspection the reported event of no image was confirmed due to defective printed circuit boards.It was also noted there is intermittent noise and dead pixels due to defective cable, shaky image due to damage coupler.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported to olympus their hd autoclavable camera head did not display an image and will not capture a picture.There was no reports of patient death, harm, or injuries related to this event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.It has been nearly 8 years since the subject device was manufactured.Based on the results of the investigation, the specific root cause of the faulty cable could not be determined at this time.The following information is stated in the instructions for use which may have prevented the event: "do not coil the camera cable with a diameter of less than 20 cm.The camera cable may be damaged.Never excessively pull the camera cable, but straighten it gradually when it is coiled.The camera cable could be damaged.Never excessively bend, pull, twist, coil, squeeze, or crush the camera cable.The camera cable could be damaged.Do not use excessive force when wiping the external surfaces of the camera cable.The camera cable could be damaged.Never attempt to lift the entire assembly by the camera cable while the camera head is attached to the endoscope.The camera cable could be damaged.Never use a clamp or forceps to attach the camera cable to another object.The camera cable could be damaged." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD AUTOCLAVABLE CAMERA HEAD
Common Device Name
AUTOCLAVABLE CAMERA HEAD
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13720118
Report Number8010047-2022-04221
Device Sequence Number19989459
Product Code FET
UDI-Device Identifier04953170310782
UDI-Public04953170310782
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K102059
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCH-S190-XZ-Q
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/20/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-