| Model Number |
98814-65 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned, extended investigation is pending at this time.A follow- up report will be submitted once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 144 mg/dl, 47 mg/dl and 317 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 144 mg/dl, 47 mg/dl and 317 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the exceed meter was reviewed and the dhrs showed the xceed meter passed all tests prior to release.Dhrs for the precision strips were reviewed and the dhrs showed the precision strips passed all tests prior to release.Retain testing was performed for precision strips and all units performed in specification and passed.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 144 mg/dl, 47 mg/dl and 317 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Meter xegz135-p078f has been returned and investigated with retained test strips.Visual inspection has been performed on the returned meter and no issues were observed.Meter powered on with button and with strip insertion.Control solution testing was performed and no issues were observed.All results were within range specification.No malfunction or product deficiency was identified.If the reported test strips are returned, an investigation will be performed.
|
| |
|