| Model Number |
98814-65 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 280 mg/dl and 38 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
No product has been returned, extended investigation is pending at this time.A follow- up report will be submitted once additional information is obtained.The device manufacturer date for the reported meter is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.The useful life of exceed meter is 4 years.As the manufacturing date of this product is before 2009, it has been in distribution beyond its useful life at the time of the complaint.Since the product exceeded its useful life, it is determined to have met specification when the product was released and through its lifespan.No further investigation activities are being performed at this time.Dhrs for the precision strips were reviewed and the dhrs showed the precision strips passed all tests prior to release.Retain testing was performed for precision strips and all units performed in specification and passed.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 280 mg/dl and 38 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 280 mg/dl and 38 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Reader (b)(6) has been returned and investigated with retained test strips.Performed visual inspection on received meter and no issues were observed.Meter powered on with button depression and test strip insertion.Control solution testing was performed and no issues were observed.All results were within range specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|