• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. STENTBOOST LIVE; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. STENTBOOST LIVE; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number STENTBOOST LIVE
Medical Device Problem Code Patient Data Problem (3197)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It has been reported to philips that when the customer was using stentboost live during a coronary procedure, the information of a previous patient was shown instead of the current patient.A reboot of the azurion system resolved the issue.No harm to the patient has been reported.Philips has started an investigation of this complaint.This complaint is reported as the issue reported relates to medical device correction 2021-igt-bst-027.
 
Additional Manufacturer Narrative
Add.Narrative: stentboost live is a software medical device that enhances visualization of stents in coronary vessels.It supports the physician in placing and deploying stents and is used in combination with a philips interventional x-ray system.Due to a software defect, the philips stentboost live r2.0 might not process x-ray images for the current run but instead show an image processed in a previous run.The image shown could be from the same patient but it could also be from another patient.The correct patient data is shown on the azurion system.Philips initiated a medical device correction (2021-igt-bst-027).On march 15, 2022, the philips service engineer implemented a software update to correct the system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STENTBOOST LIVE
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
veenpluis 4-6
best 5684 -PC
NL   5684 PC
MDR Report Key13720680
Report Number3003768277-2022-00104
Device Sequence Number4807409
Product Code OWB
UDI-Device Identifier00884838101890
UDI-Public00884838101890
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K170144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSTENTBOOST LIVE
Device Catalogue Number001017
Device Lot Number(10)2.0
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer02/24/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-