| Model Number |
2545-00-711 |
| Medical Device Problem Codes |
Break (1069); Illegible Information (4050)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, it was confirmed that the product code of implant trial is illegible.It is unused.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received states that the breakage of the instrument is not reported but the part number is missing and unreadable.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: the device associated with this report was returned to depuy synthes for evaluation.Visual examination found, the product code numbers partially imprinted on the device.Confirming, the reported allegation.However, no broken condition could be observed.Additional monitoring, for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a search of the depuy nonconformance (nc) quality system found no past nc¿s associated with this product/lot combination.The evidence so far shows, that this is an isolated case with no further reports regarding this lot#: mvmgbg390.An nc# was created to address this current complaint issue.
|
| |
|