• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD¿ SHARPS COLLECTOR; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD¿ SHARPS COLLECTOR; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305616
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the slide portion of the bd¿ sharps collector lid was missing.The following information was provided by the initial reporter, translated from japanese: "the customer's verbatim report is that the slide part of the lid was missing.".
 
Additional Manufacturer Narrative
Oem manufacturer: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Medical device expiration date: na.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the slide portion of the bd¿ sharps collector lid was missing.The following information was provided by the initial reporter, translated from japanese: "the customer's verbatim report is that the slide part of the lid was missing.".
 
Additional Manufacturer Narrative
Investigation summary: according to the dhr review process; the result showed there were no issues reported like missing slider during the manufacturing process of the part number reported under this customer complaint.A review of the ncmr¿s was performed; the result showed there were no issues reported like missing slider for the same part number throughout the last twelve months.Based on this investigation it was determined that this failure mode is related to the manufacturing process since the lack of slider would be an omission error within process.Due to this failure mode was already known in the process, an improvement action (vision system), had already been implemented in the main.According to risk assessment result, this issue was considered low risk therefore a corrective action isn¿t needed, also a quality alert was posted within the process in order to make aware all the people involved in the manufacturing process of this product.All the complaint information was captured also for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ SHARPS COLLECTOR
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13720744
Report Number2243072-2022-00328
Device Sequence Number14576451
Product Code FMI
UDI-Device Identifier00382903056163
UDI-Public00382903056163
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K943134
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number305616
Device Catalogue Number305616
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer05/10/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date05/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-