| Model Number |
MS1-4570S |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that the rod failed to distract.No patient adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned and is pending device evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: upon receipt of the returned product score marks consistent with incremental distraction on the distraction rod.The distraction rod was able to be lengthened by hand which indicated a broken locking pin.X-ray images of the internal components and sectioning the rod confirmed the locking pin was broken.Sectioning the rod additionally revealed corrosion around the looking pin.As a result of the locking pin breakage, there is nothing constraining the axial motion of the distraction rod within the housing tube; this is usually known as "pistoning" which confirmed the reported failure.The root cause is likely due to corrosion and bending stresses.A capa was completed in which a new pin was implemented with 65 percent improved strength and significantly improved corrosion resistance.Device history records (dhr) review revealed that this unit was built prior to the release of the new pin.Device records review: a review of the dhr confirmed the unit met all of the required quality inspections and specifications prior to release.
|
| |
|
Search Alerts/Recalls
|