• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® HANDPIECE - MAGNUM® II; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® HANDPIECE - MAGNUM® II; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I Back to Search Results
Model Number 1897200
Medical Device Problem Codes Overheating of Device (1437); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: unable to perform temperature test as blade was not rotating.Visual inspection found the handpiece to be cosmetically not ok.Connector has dent, hi-pot test failed and the blade/bur was not rotating.The motor cable assembly was found faulty.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare professional (hcp) reported that the handpiece was heating up and the blade was not rotating.Event found pre-op.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® HANDPIECE - MAGNUM® II
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key13721130
Report Number1045254-2022-00133
Device Sequence Number4494646
Product Code ERL
UDI-Device Identifier00643169360846
UDI-Public00643169360846
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K002224
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1897200
Device Catalogue Number1897200
Device Lot Number216719002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2022
Initial Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-