| Model Number |
1897200 |
| Medical Device Problem Codes |
Overheating of Device (1437); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: unable to perform temperature test as blade was not rotating.Visual inspection found the handpiece to be cosmetically not ok.Connector has dent, hi-pot test failed and the blade/bur was not rotating.The motor cable assembly was found faulty.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A healthcare professional (hcp) reported that the handpiece was heating up and the blade was not rotating.Event found pre-op.There was no patient involvement.
|
| |
|