DAVIS & GECK CARIBE LTD VERSAONE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number NONB15STF
Medical Device Problem Codes Difficult to Insert (1316); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic duodenal switch, the seal became increasingly difficult to pass the stapler through it.The device was sticky and grabbing the smooth shaft of the stapler making it difficult to slide the stapler.Surgeon had to use extra force to push the device into position.This happened with multiple firing of stapler.Each use needed an increase force to pass the instrument.Additionally, the seal was leaking gas.The surgeon had to replaced it with a new one to complete the procedure.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the obturator was received inserted into the cannula, prolonged insertion can cause the duckbill seal to become stretched.The trocar and cannula appeared intact.The circular seal and duckbill seal were visibly stretched out.No information regarding the specific customer issue that occurred with the device was available.It was reported that the user experienced difficulty in the insertion or removal of the device to or from the port and the device was leaking fluid or air.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAONE
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13721142
Report Number9612501-2022-00385
Device Sequence Number17266786
Product Code GCJ
UDI-Device Identifier10884521548091
UDI-Public10884521548091
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K160230
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model NumberNONB15STF
Device Catalogue NumberNONB15STF
Device Lot NumberJ1F2376Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer04/22/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-