|
It was reported that a patient underwent a c-section procedure on (b)(6) 2022 and suture was used.During the procedure, when applying tension on the suture to approximate the tissue, the needle would dislodge at the swage right where the suture attaches to the needle.This happened 4 times.There were no patient consequences reported.Additional information was requested.
|
|
(b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained via: was the needle it retrieved during the same surgical procedure? yes.What tissue/location was suturing when pull off occurred? fascia abdominal closure for c section.Were there any unexpected outcomes or complications as a result of this suture needle pull off event? no.Patient condition? no patient consequence.Procedure date? (b)(6) 2022.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Events reported via: 2210968-2022-01657, 2210968-2022-01658 and 2210968-2022-01659.
|
|
Product complaint # (b)(4).Date sent to the fda: 4/11/2022 h6 component code: g07002 no device problem found.Additional information: d4, d9, h3, h4, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: one needle/suture piece of product code j977h was returned to ethicon inc for analysis.In order to evaluate the conditions of the returned sample, the swage and attachment area was noted to be as expected and a suture piece still was attached to the barrel hole needle and the suture end was cut by a sharp edge.No needle pull off was observed.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.Functional test was not performed, due to the condition of the sample received.The manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: three opened boxes with thirty- five, seventeen and twenty-seven packets ea.Of product code j977h were returned to ethicon inc for analysis with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strands to detect any issue related to no defects were observed during evaluation.Functional test was performed, and the pull force meet the requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Events reported via:2210968-2022-01657, and 2210968-2022-01659.
|