• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD UD 36IN 1 S/A CTX; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD UD 36IN 1 S/A CTX; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number J977H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained via: was the needle it retrieved during the same surgical procedure? yes.What tissue/location was suturing when pull off occurred? fascia abdominal closure for c section.3) were there any unexpected outcomes or complications as a result of this suture needle pull off event? no.Patient condition? no patient consequence.Procedure date? (b)(6) 2022.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Events reported via: 2210968-2022-01656, 2210968-2022-01658 and 2210968-2022-01659.
 
Event or Problem Description
It was reported that a patient underwent a c-section procedure on (b)(6) 2022 and suture was used.During the procedure, when applying tension on the suture to approximate the tissue, the needle would dislodge at the swage right where the suture attaches to the needle.This happened 4 times.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 4/11/2022.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: one needle/suture piece of product code j977h was returned to ethicon inc for analysis.In order to evaluate the conditions of the returned sample, the swage and attachment area was noted to be as expected and a suture piece still was attached to the barrel hole needle and the suture end was cut by a sharp edge.No needle pull off was observed.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.Functional test was not performed, due to the condition of the sample received.The manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: three opened boxes with thirty- five, seventeen and twenty-seven packets ea.Of product code j977h were returned to ethicon inc for analysis with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strands to detect any issue related to no defects were observed during evaluation.Functional test was performed, and the pull force meet the requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Events reported via: 2210968-2022-01656, 2210968-2022-01657, and 2210968-2022-01659.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD UD 36IN 1 S/A CTX
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13721834
Report Number2210968-2022-01657
Device Sequence Number4806393
Product Code GAM
UDI-Device Identifier10705031043794
UDI-Public10705031043794
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model NumberJ977H
Device Catalogue NumberJ977H
Device Lot NumberRDMCXJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2022
Supplement Date Received by Manufacturer03/21/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-