Model Number 7134
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft fracture occurred.A 2.50mm x 20mm emerge balloon catheter was advanced for dilatation.However, during insertion, the device fractured at the middle of the shaft.The device was removed and the procedure was completed with a different device.No patient complications were reported.
 
Event or Problem Description
It was reported that shaft fracture occurred.A 2.50mm x 20mm emerge balloon catheter was advanced for dilatation.However, during insertion, the device fractured at the middle of the shaft.The device was removed and the procedure was completed with a different device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an emerge mr balloon catheter.The device was visually and microscopically examined.There were numerous kinks to the hypotube of the device.There was a separation of the hypotube 66.5cm from the strain relief.The separated ends of the hypotube were ovaled in shape indicating the device was kinked prior to separation.There was contrast in the inflation lumen and the balloon was tightly folded.Photo media reviewed depicts a separation and kink in the hypotube of the device which aligns to some of the damages found in analysis.Product analysis confirmed the reported break as there was separation to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13721891
Report Number2134265-2022-02654
Device Sequence Number14556609
Product Code LOX
UDI-Device Identifier08714729806325
UDI-Public08714729806325
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/19/2024
Device Model Number7134
Device Catalogue Number7134
Device Lot Number0028607192
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-