• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OMNIGRAFT DERMAL REGENERATION MATRIX 2.5CM X 2.5CM; INTEGRA OMNIGRAFT DERMAL REGENERATION MATRIX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OMNIGRAFT DERMAL REGENERATION MATRIX 2.5CM X 2.5CM; INTEGRA OMNIGRAFT DERMAL REGENERATION MATRIX Back to Search Results
Model Number DFU25251
Medical Device Problem Code Expiration Date Error (2528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Omnigraft dermal (dfu25251) was not returned for evaluation.Device history record (dhr) review - no anomalies found.The label for lot 5329553 is applicable to the product packaging, while the label for lot 4742283 is applicable to the omnigraft device.Omnigraft expiration date is 36 months from the month of packaging (the first day the material was packaged into the primary packaging) which for this lot was 11jun2020.It is confirmed that the lowest expiration date was used from the component items to obtain the overall device lot number: 30jun2023.Use of the earliest expiration date is a conservative approach taken to ensure the entire contents do not get used past the earliest expiration date of any of the finished good¿s components.The product was within its expiry at the time of patient contact on 10nov2021.Failure analysis - the investigation confirms that there is no basis to the complaint.The presence of the labels was appropriate and that the expiration dates and lot numbers were correct.Root cause - the complaint was predicated on the complainant¿s concern over the product¿s dissimilar lot numbers and expiration date.This investigation confirmed that the presence of those labels was appropriate and that the expiration dates printed on them were correct.Therefore, there is no root cause to determine as there is no basis to the complaint.
 
Event or Problem Description
Medwatch report number: (b)(4).A facility reported that the outer package of the omnigraft dermal (dfu25251) has different lot number and expiration date than the inner package.Outer expiration: 2022-01-05; while inner expiration: 2023-06-30.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIGRAFT DERMAL REGENERATION MATRIX 2.5CM X 2.5CM
Common Device Name
INTEGRA OMNIGRAFT DERMAL REGENERATION MATRIX
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION
105 morgan lane
105 morgan lane
plainsboro NJ
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION
105 morgan lane
plainsboro NJ
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13722107
Report Number1121308-2022-00012
Device Sequence Number14608363
Product Code MDD
UDI-Device Identifier10381780258872
UDI-Public10381780258872
Combination Product (Y/N)N
PMA/510(K) Number
P900033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDFU25251
Device Catalogue NumberDFU25251
Device Lot Number5329553, 442283
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/10/2022
Initial Report FDA Received Date03/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age52 YR
Patient SexUnknown
-
-