Catalog Number 955467
Device Problems Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2022
Event Type  malfunction  
Event Description
It was reported that during use with a prismaflex m150 set, an external fluid leak was observed from the pump segment due to a cut hole of the tube.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Initial reporter facility name: the first affiliated hospital of (b)(6) university should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the actual sample was not received for investigation however photos were provided.The visual inspection observed that the replacement pump segment was perforated near the support plate.The reported condition was verified.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M150 SET
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13722187
MDR Text Key287184891
Report Number8010182-2022-00057
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414123451
UDI-Public(01)07332414123451
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Catalogue Number955467
Device Lot Number21J0501CA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/31/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-