• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INFUTRONIX, LLC. ADMINISTRATION SET FOR HALO II AMBULATORY INFUSION PUMP; IV ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INFUTRONIX, LLC. ADMINISTRATION SET FOR HALO II AMBULATORY INFUSION PUMP; IV ADMINISTRATION SET Back to Search Results
Model Number HS-004
Medical Device Problem Codes Loss of or Failure to Bond (1068); Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On (b)(6) 2022, a service provider of infutronix provided the image for the affected administration set and visual inspection confirmed that the set had disconnected from the proximal end of the cassette connector.The reported issue was confirmed.The affected device was never returned for evaluation and therefore no root cause identified.
 
Event or Problem Description
On (b)(6) 2022, infutronix received a complaint from an end user: "an administration set model hs-004, lot 2106005 set broke at the connection of the cassette." device operator was a registered nurse.Medication infused was 5fu.A patient was involved but not harmed.The contract manufacturer of the affected device is (b)(6) medical co.Ltd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADMINISTRATION SET FOR HALO II AMBULATORY INFUSION PUMP
Common Device Name
IV ADMINISTRATION SET
Manufacturer (Section D)
INFUTRONIX, LLC.
177 pine street
natick MA 01760
Manufacturer (Section G)
INFUTRONIX, LLC
177 pine street
natick MA 01760
Manufacturer Contact
chaoyoung lee
177 pine street
natick, MA 01760
5086502007
MDR Report Key13722877
Report Number3011581906-2022-00031
Device Sequence Number14630262
Product Code FPA
UDI-Device Identifier00817170020031
UDI-Public00817170020031
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K153193
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2024
Device Model NumberHS-004
Device Catalogue NumberHS-004
Device Lot Number2106005
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Initial Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-