| Model Number |
VITAL SIGNS® PRESSURE INFUSOR |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the vital signs® pressure infusor air output is draining gradually.The issue occurred during patient-use for 72 hours.As of this time, there is no information regarding patient harm with this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation update: d9, g3, g6, h2, h3, h6 and h10.Results of investigation: the suspect device was evaluated and tested.Vyaire medical was unable to confirmed the reported issue.Based on the results of the leakage test instructions mwi-51004353-as-009/i (pressure leakage test) and mwi-51004353-as-010/g (leakage test 3 hours), we can generally determine if the leak occurred during the manufacturing process.The return sample met the air leakage standard.According to the dhr, all units were produced using the mcp-51004353-as-001/f production flow.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|