• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; UROLOGICAL CATHETER AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; UROLOGICAL CATHETER AND ACCESSORIES Back to Search Results
Model Number M006630222B1
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not received for evaluation at the time of this report; therefore, no physical analysis of the device can be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at (b)(6) medical and was determined to be acceptable.As noted in the device instructions for use (dfu) warnings: do not manipulate, advance and/or withdraw the zipwire hydrophilic guidewire through a metal cannula or needle.Manipulation, advancement, and/or withdrawal through a metal device may result in destruction and/or separation of the outer polymer jacket requiring retrieval.If a needle is used for initial placement, a plastic entry needle is recommended when using the zipwire hydrophilic guidewire.Extreme caution should be observed when used with a one-wall puncture needle.Manipulate the zipwire hydrophilic guidewire slowly and carefully in the urinary system while confirming the behavior and location of the wire's tip under fluoroscopy.Excessive manipulation of the zipwire hydrophilic guidewire without fluoroscopic confirmation may result in perforation or trauma of the linings or associated tissues, channels or ducts.If any resistance is felt or if the tip's behavior and/or location seem improper, stop manipulating the wire and/or catheter and determine the cause by fluoroscopy.Failure to exercise proper caution may result in bending, kinking, separation of the zipwire hydrophilic guidewire's tip, damage to the catheter, or damage to the urinary system.If necessary, remove the zipwire hydrophilic guidewire and ancillary device or scope as a complete unit to avoid complications.The dfu precautions also indicate: the zipwire hydrophilic guidewire should be advanced through the scope using short, deliberate 2-3cm movements to prevent inadvertent damage to the device or patient.Due to variations in certain catheter tip diameters, abrasion of the hydrophilic coating may occur during manipulation.If any resistance is felt during introduction of the catheter, it is advisable to stop using such catheters.The complaint is non-verifiable as the product was not returned for evaluation.At this time, it is not possible to assign a definitive root cause for the event as reported.If any further information is received, a follow up medwatch report will be submitted.
 
Event or Problem Description
Event description: it was reported that: used during a percutaneous nephrostomy, the distal end of the guide deteriorated into mini-filaments, a piece of which remained inside the patient's kidney (visible on the x-ray).Clinical consequences: a guidewire filament remained inside the patient's kidney.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIPWIRE HYDROPHILIC GUIDEWIRE
Common Device Name
UROLOGICAL CATHETER AND ACCESSORIES
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer Contact
victoria foss
340 lake hazeltine drive
chaska, MN 55318
9526418506
MDR Report Key13724938
Report Number2126666-2022-00012
Device Sequence Number17562418
Product Code EYA
UDI-Device Identifier08714729755326
UDI-Public08714729755326
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Pharmacist
Type of Report Initial
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model NumberM006630222B1
Device Catalogue Number630-222B
Device Lot Number6378769
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2022
Initial Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-