• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 600 ML, 2-14 ML/HR; ELASTOMERIC - SAF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 600 ML, 2-14 ML/HR; ELASTOMERIC - SAF Back to Search Results
Model Number CB6004
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Pain (1994)
Date of Event 02/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
Fill volume: 600 ml.Flow rate: 8ml/hr.Procedure: unknown.Cathplace: unknown.Infusion start time: unknown.Infusion stop time: unknown.Drug: 0.2% ropivacaine.It was reported that the pump emptied 24 hours sooner than expected.The patient experienced pain after the pump was emptied and was given oxy/acetaminophen more often than anticipated.Additional information has been requested but not yet received.
 
Additional Manufacturer Narrative
The sample is reported to be available but has not yet been received by the manufacturer.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 09 mar 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).
 
Additional Manufacturer Narrative
The product involved in the report has been returned and the investigation remains in progress at this time.All information reasonably known as of 22 mar 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
Per additional information received 14 mar 2022: procedure: left acl repair.Cathplace: tuohy 9cm.Infusion start time: (b)(6) 2022 between noon and 1pm.Infusion stop time: (b)(6) 2022 in the evening, exact time unknown.
 
Additional Manufacturer Narrative
One used pump with extension set and catheter was received for evaluation.Flow testing of the various flow rates via the select-a-flow (saf) revealed that the pump was working within specification.The complaint could not be confirmed as reported.No root cause could be determined.The device history record for lot 20066764 was reviewed and the product was produced according to product specifications.All information reasonably known as of 11 apr 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 600 ML, 2-14 ML/HR
Common Device Name
ELASTOMERIC - SAF
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c.
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key13725125
Report Number2026095-2022-00012
Device Sequence Number12233013
Product Code MEB
UDI-Device Identifier00193494134761
UDI-Public00193494134761
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K063530
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/13/2023
Device Model NumberCB6004
Device Catalogue NumberN/A
Device Lot Number20066764
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2022
Supplement Date Received by Manufacturer03/12/2022
04/11/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date03/23/2022
04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age16 YR
Patient SexMale
Patient Weight54 KG
-
-