| Model Number |
INS-0352 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Physician experienced a brush sheath detaching from the brush while deploying a fiducial during a procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
3007222345-2022-00011 occured during the same procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide information that was inadvertently not included in the initial medwatch report and to provide the investigation results.G3: date received by manufacturer was missing from the initial report and should have been 11-feb-2022.The device was not returned to veran for investigation.Based on the results of the investigation, the complaint data indicates that the ins-0352 broke within the sheath while placing a fiducial.There will be a sheath anchoring and adhesion design change to address this issue.Veran will continue to monitor field performance for this device.This supplemental report is being submitted as a corrective action following a retrospective review of complaints in response to an observation from an external inspection.
|
| |
|
Event or Problem Description
|
|
Additional information regarding the reported event: there was no impact on the patient or the procedure due to the reported issue.The procedure was completed with no delay after switching to a new brush.
|
| |
|
Search Alerts/Recalls
|