VERAN MEDICAL TECHNOLOGIES, INC SPIN ACCESS CATHETER 90 DEGREES Back to Search Results
Model Number INS-5910
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Doctor has a spin access catheter 90 degree handle break off while attempting to turn the catheter.There was a delay in procedure for 30 minutes.The procedure was aborted.
 
Additional Manufacturer Narrative
This report is being supplemented to provide information that was inadvertently not provided in the initial medwatch report, to correct information provided in the initial report, and to provide the results of the investigation of the event.Correction to b1, b2, and h1: this event is considered to be an adverse event/serious injury due to the 30 minute procedure delay and procedure being discontinued after the patient was under anesthesia.G3 of the initial medwatch report was blank.The date received by manufacturer was 10-mar-2022.A full investigation could not be completed due to the device not being returned for evaluation, as it was discarded by the facility.The investigation found that the drawing for the device does not provide specifications for handle attachment.There are no specific inspection requirements for the handle.A root cause of the reported event could not be determined.Veran will continue to monitor field performance for this device.This supplemental report is being submitted as a corrective action following a retrospective review of complaints in response to an observation from an external inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIN ACCESS CATHETER 90 DEGREES
Common Device Name
CATHETER
Manufacturer (Section D)
VERAN MEDICAL TECHNOLOGIES, INC
1938 innerbelt business center
saint louis MO 63114
Manufacturer (Section G)
VERAN MEDICAL TECHNOLOGIES, INC
1938 innerbelt business center
saint louis MO 63114
Manufacturer Contact
jennifer andre
1938 innerbelt business center
saint louis, MO 63114
3146598500
MDR Report Key13725570
Report Number3007222345-2022-00009
Device Sequence Number4025499
Product Code JAK
UDI-Device Identifier00815686021276
UDI-Public00815686021276
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K170023
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/10/2023
Device Model NumberINS-5910
Device Catalogue NumberINS-5910
Device Lot Number85028200116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/10/2023
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date08/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-