| Model Number |
7213 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.The 85% stenosed target lesion was located in a mildly tortuous and severely calcified proximal right coronary artery.A 3.00mm x 15mm nc emerge balloon catheter was advanced for dilation.While the device was being pushing in a tight calcified lesion, about 1/4th of the proximal shaft fractured.The device was removed together with the wire and the procedure completed with another emerge balloon catheter.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.The 85% stenosed target lesion was located in a mildly tortuous and severely calcified proximal right coronary artery.A 3.00mm x 15mm nc emerge balloon catheter was advanced for dilation.While the device was being pushing in a tight calcified lesion, about 1/4th of the proximal shaft fractured.The device was removed together with the wire and the procedure completed with another emerge balloon catheter.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned product consisted of an nc emerge catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.Visual examination confirmed a break.Microscopic examination revealed kinks 7.03 cm from the hub and on the separated piece at 24.2 cm and 2.6 cm from the distal end.
|
| |
|