• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC TLS 5F MAXIMAL BARRIER DL W/MICROINTRODUCER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC TLS 5F MAXIMAL BARRIER DL W/MICROINTRODUCER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of refx4493 showed seven other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "two piccs that have leaked where the hub meets the extension set." no other information was provided.This report addresses the second device.
 
Event or Problem Description
It was reported "two piccs that have leaked where the hub meets the extension set." no other information was provided.This report addresses the second device.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of leaking catheters was confirmed and the cause was manufacturing-related.The product returned for evaluation was two 5fr d/l powerpicc catheters.Light usage residues were observed throughout both samples and needleless injection caps were attached to all four luer adapters.Attempts to infuse water through the samples using a 12ml syringe revealed all four lumens to be patent to infusion; however, leaks were observed at both purple luer/injection cap joints.Inspection of both purple luer adapters using an iso taper gauge revealed that the gauge could not be fully inserted.Microscopic inspection of the purple luer adapters revealed protrusions in the luer material within the orifices.The mold defects appeared to prevent firm attachment of the injection caps, as well as full insertion of the taper gauge.Longitudinally aligned scoring marks were observed within the orifices adjacent to the mold defect.The molding defect observed within the luer orifices prevented firm attachment with the injection caps and resulted in leakage at the interface between the catheter and the cap.Photographs of the sample have been forwarded to the manufacturing site for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC TLS 5F MAXIMAL BARRIER DL W/MICROINTRODUCER
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13726069
Report Number3006260740-2022-00668
Device Sequence Number6055131
Product Code LJS
UDI-Device Identifier00801741035692
UDI-Public(01)00801741035692
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K063240
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number9275108D
Device Lot NumberREFX4493
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer05/06/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date05/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-