| Model Number |
35615 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: an introducer sheath, and sheath dilator and unknown guidewire was returned for analysis.Visual inspection revealed that the introducer sheath and sheath dilator were observed that the distal tip was torn.The unknown guidewire was observed damage at distal tip.No other issues or damage were observed.
|
| |
|
Event or Problem Description
|
|
Reportable based on device investigation completed on 17feb2022.It was reported that the introducer wire got unraveled.The target lesion was located in the kidney.A.038 accustick ii system was selected for use.During the procedure, it was noted that the introducer wire was caught on the introducer and got unraveled in the process of getting out of the introducer.The procedure was completed with another of the same device.There were no complications reported.However, device investigations revealed that the introducer sheath and sheath dilator were observed torn.
|
| |
|
Search Alerts/Recalls
|