Model Number 35615
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by manufacturer: an introducer sheath, and sheath dilator and unknown guidewire was returned for analysis.Visual inspection revealed that the introducer sheath and sheath dilator were observed that the distal tip was torn.The unknown guidewire was observed damage at distal tip.No other issues or damage were observed.
 
Event or Problem Description
Reportable based on device investigation completed on 17feb2022.It was reported that the introducer wire got unraveled.The target lesion was located in the kidney.A.038 accustick ii system was selected for use.During the procedure, it was noted that the introducer wire was caught on the introducer and got unraveled in the process of getting out of the introducer.The procedure was completed with another of the same device.There were no complications reported.However, device investigations revealed that the introducer sheath and sheath dilator were observed torn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUSTICK II
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13726922
Report Number2134265-2022-02548
Device Sequence Number5882945
Product Code KGZ
UDI-Device Identifier08714729157601
UDI-Public08714729157601
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/05/2024
Device Model Number35615
Device Catalogue Number35615
Device Lot Number0028150786
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2022
Initial Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-