| Model Number |
1012450-15 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a moderately calcified, moderately tortuous, 90% stenosed lesion in the mid-right coronary artery (rca).A 3.25x15mm nc trek balloon dilation catheter (bdc) was advanced to the target lesion for pre-dilatation and was inflated 1 time to 10 atmospheres (atm) for 10 seconds, when a balloon rupture occurred.The bdc was removed without issue.There was no adverse patient effect and no clinically significant delay in the procedure.A non-abbott device was used to pre-dilatate, after which a 3.5x18mm xience skypoint drug eluting stent (des) was implanted to successfully complete the procedure.No additional information was provided.
|
| |
|