| Model Number |
DIU150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
11/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The exact date is unknown, the best estimate is after the implant date (b)(6) 2021 and prior to the explant date (b)(6) 2022.The device has not returned for evaluation.Therefore, a visual analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain the missing information.However, to date it has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that after implanting an intraocular lens (iol) into the patient¿s left eye, their distance vision was not optimized, not sharp.The left was monovision for distance, had residual hyperopia and astigmatism.Reportedly, in november, the patient¿s vision was not optimal.The iol was explanted (b)(6) 2022.There was no incision enlargement, sutures, or vitrectomy required.Pre-operative visual acuity 20/40 and post operative visual acuity is 20/25- 20/30.The iol exchange was successful as the final outcome improved following the iol explant.On (b)(6) 2022, the patient¿s visual acuity was 20/25 sc (without correction) and best corrected 20/20-.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: additional information received indicated that the patient was white/american with weight of 155 pounds.The date of first symptoms was noted to be (b)(6) 2021.It was also indicated that the replacement intraocular lens (iol) was a 17.5 diopter iol of the same model.The following fields were updated accordingly: section a4: patient weight: 155 pounds.Section a5: race: white/american.Section b3: date of event: (b)(6), 2021.Corrected data: upon further review, it was noticed that since the original lens was 16.5 diopters and the replacement lens was 17.5 diopter, there is a significant change of 1.0 diopter which indicates that the patient would have needed a different lens power to be calculated/used.As the reported event was not due to the lens product problem or malfunction.Therefore, the event is no longer reportable and no further information will be provided under this manufacturer report number 3012236936-2022-00610.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|
| |
|
Search Alerts/Recalls
|
|
|