AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DIB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that upon implanting the intraocular lens (iol), the doctor noticed a crack in the body of the lens.The iol was removed and another johnson & johnson lens (same model and diopter) was implanted as a replacement.The incision was slightly enlarged.The iol was discarded and will not be returned to j&j.Reportedly, the patient is doing well.No further information was provided.
 
Additional Manufacturer Narrative
The intraocular lens (iol) is not returning for evaluation as it is discarded; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain the missing information.However, to date a response has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13727421
Report Number3012236936-2022-00587
Device Sequence Number11963830
Product Code HQL
UDI-Device Identifier05050474731622
UDI-Public(01)05050474731622(17)241004
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/04/2024
Device Model NumberDIB00
Device Catalogue NumberDIB00U0135
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2022
Initial Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-