• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL ADVANCED LOCKING SCREW IMN SCREWS Ø5X35MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL ADVANCED LOCKING SCREW IMN SCREWS Ø5X35MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 2361-5035S
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
 
Event or Problem Description
It was reported: "during ap screw insertion to distal of the tibia, the screw became stuck in the nail.An attempt was made to remove it, but it was difficult and took 30 minutes.The surgery was completed without inserting any screws into the hole.".
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidence was provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Finally, the use of a locking screw is mandatory and the ifu points out as following ¿the advanced locking screw must be inserted using reasonable force.To provide desired function and to avoid damaging the screw.If unreasonable insertion torque is noticed, stop insertion, turn the screw counterclockwise and then attempt to insert the screw.If unreasonable insertion torque is still noticed, remove the screw and proceed with a locking screw.¿ in case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
 
Event or Problem Description
It was reported: "during ap screw insertion to distal of the tibia, the screw became stuck in the nail.An attempt was made to remove it, but it was difficult and took 30 minutes.The surgery was completed without inserting any screws into the hole.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED LOCKING SCREW IMN SCREWS Ø5X35MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13727591
Report Number0009610622-2022-00087
Device Sequence Number4623363
Product Code HWC
UDI-Device Identifier07613327381597
UDI-Public07613327381597
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K191271
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2361-5035S
Device Catalogue Number23615035S
Device Lot NumberK0459DC
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/10/2022
Supplement Date Received by Manufacturer05/12/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-