The reported event could be confirmed.The device inspection revealed the following: the product returned is broken in two at the level of the 7th screw hole.Signs of a fatigue fracture are visible with the microscopic analysis of the broken area.This gives evidence that the fracture was caused by repeated overloading of the plate.Based on investigation, the root cause was attributed to a patient related issue.The failure was probably caused by repeated overloading of the plate.A probable poor bone condition, related to the advanced age of the patient (91 years old) is likely to have influenced a poor recovery and the fracture of the device.It was reported that the patient ¿apparently did dance¿, thus the patient was active and did not fully respect the recovery time advised (the reported fracture occurred 2 months after implantation).This supports the conclusion that the fracture was caused by fatigue of the plate.In order to have a more complete conclusion as of the patient¿s bone condition, x-rays of the fracture would be required.As a reminder, the ifu clearly states: " the surgeon must warn the patient that the device cannot and does not replicate a normal healthy bone, that the device can break or become damaged as a result of strenuous activity, trauma, mal-union or nonunion and that the device has a finite expected service life and may need to be removed at some time in the future.[.] these implants are neither intended to carry the full load of the patient acutely, nor intended to carry a significant portion of the load for extended periods of time.For this reason post-operative instructions and warnings to patients are extremely important.External immobilization (e.G.Bracing or casting) may be employed until x-rays or other procedures confirm adequate bone consolidation." a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
|