Model Number 7590
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the severely tortuous and severely calcified middle left anterior descending artery.A 2.50mm x 20mm maverick balloon catheter was advanced for dilatation.The balloon was inflated twice at 12 atmospheres respectively for 21-22 seconds and the balloon ruptured.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a maverick 2mr balloon catheter.The device was visually and microscopically examined.At 4mm and 2.6cm from the strain relief, the hypotube of the device was kinked.There was contrast in the inflation lumen and balloon.There was blood in the guidewire lumen and the balloon was loosely folded.There was tip damage to the device.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.The device was able to inflate to rated burst pressure and deflate with no issue.Product analysis could not confirm reported event as the device inflated to rated burst pressure and deflated with no issue.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the severely tortuous and severely calcified middle left anterior descending artery.A 2.50mm x 20mm maverick balloon catheter was advanced for dilatation.The balloon was inflated twice at 12 atmospheres respectively for 21-22 seconds and the balloon ruptured.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13728166
Report Number2134265-2022-02770
Device Sequence Number17266734
Product Code LOX
UDI-Device Identifier08714729370185
UDI-Public08714729370185
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/25/2024
Device Model Number7590
Device Catalogue Number7590
Device Lot Number0027385598
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date03/10/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
-
-