• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. WHITESTAR SIGNATURE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. WHITESTAR SIGNATURE; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number OPO73
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a foreign matter, which looked like a piece of metal 2-3mm long, entered into patient¿s eye at the beginning of phaco procedure during surgery.The foreign matter was immediately removed from patient¿s eye.There was no patient injury.
 
Additional Manufacturer Narrative
Weight is unknown/no information has been provided.At the time of the investigation, product was not available for evaluation, therefore no testing could be performed.The reported event cannot be confirmed.The review of the device history record (dhr) for the patient interface showed that there were no issues or non-conformities.The device and its components met all specifications prior to being released.Manufacturing has been ruled out as a potential cause for the reported issue.Based on the investigation results, no corrective action has been issued.Based on the investigation results there is no indication of a product quality deficiency.Johnson & johnson surgical vision will continue to monitor this type of complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WHITESTAR SIGNATURE
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
31 technology drive
irvine CA 92618
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13728450
Report Number3012236936-2022-00598
Device Sequence Number4709962
Product Code HQC
UDI-Device Identifier15050474602080
UDI-Public(01)15050474602080(17)240215(10)60282074
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160236
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/15/2024
Device Model NumberOPO73
Device Catalogue NumberOPO73
Device Lot Number60282074
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2022
Initial Report FDA Received Date03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-